Theoretical and Experimental Study of the Distribution of Radiation Dose of Electron Beams Used in Superficial Tumours Treatment

Authors

  • Haisam Ibrahim Jbeli Tishreen University
  • Bassam Soleman Saad Tishreen University
  • Yara Mahmoud Mohammad Al Andalus, University
  • Manal Moein Alkobily Tishreen University

Abstract

 

In this work, a flat ionization chamber and a homogenous measurement medium were used to study the percentage changes in the depth dose of the electron beams produced by the Varian Clinic 23ix in Tishreen University Hospital in Lattakia, And study the features of the profile of these packages, which include flattening, symmetry and shadow, and then compared the experimental results with the reference values declared globally.

The accuracy of accelerator's labour used in this research is strongly confirmed since the result of the comparison refer to a perfect macth with the abovementioned values.

References

–أيهم المحمود, حساب الجرعة في المعالجة الإشعاعية عن قرب لسرطان الرحم باستعمال الكود MCNP, درجة ماجستير, جامعة دمشق, 2012.

- C.DESROSIERS, V.MOSKVIN, A.F.BIELAJEW, L.PAPIEZ, 150-250 MeV electron beams in radiation therapy,UK, Phys. Med. Biol. 45 (2000) 1781–1805.

- T.CONNELL, A.ALEXANDER, M.EYVANS, J.SEUNTJENS, An experimental feasibility study on the use of scattering foil free beams for modulated electron radiotherapy, Canada, Phys. Med. Biol. 57 (2012) 3259–3272.

- C.D.VENAZIO, M.MARINELLI, A. TONNETTI, G.V.RINATI, P.BAGALA, M.D.FALCO, A.S.GUERRA, M.PIMPINELLA, Comparison between small radiation therapy electron beams collimated by cerrobend and tubular applicators, Itally,

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1(2015) 329-335.

-[5]جبيلي, سعد, علي حسن, قياس توزع الجرعة الإشعاعية تجريبياً ومقارنتها مع حسابات نظام تخطيط الجرعة TPS , اللاذقية, مجلة جامعة تشرين للبحوث والدراسات العلمية, العدد2 (2015) 17.

- P.PATHAK, P.K.MISHRA, M.SINGH, P.K.MISHRA. Analyticl Study of flatness and Symmetry of electron beam with 2D Array Detectors. India, October ( 2015) 294-301.

-[7]محمد حمودي, قياس توزع الجرعة الإشعاعية في التطبيقات المختلفة للمعالجة بالإشعاع, درجة ماجستير, جامعة دمشق, 2012.

- E.B. PODGORSAK. Radiation Oncology Physics : A Hand book for teacher and student. Vienna, 2005.

- SHIQIN SU, K.MORAN, J.L.ROBAR, Design and Production of 3D printed Bolus for electron Radiation therapy. Dalhousie University, JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 4(2014) 194-211.

- SHIQIN SU, Design and Production of 3D printed Bolus for electron Radiation therapy, master degree, Dalhousie University, August 2014.

-International Atomic Energy Agency (IAEA), TRS, ABSORBED DOSE DETERMENATION IN EXTERNAL BEAM RADIOTHERAPY, Vienna, 2000.

- K.R.HOGSTROM, P.ALOMOND, Review of electron beam therapy physics, USA, Phys. Med. Biol. 51 (2006) R455–R489.

Published

2019-02-28

How to Cite

1.
Ibrahim Jbeli H, Soleman Saad B, Mahmoud Mohammad Y, Moein Alkobily M. Theoretical and Experimental Study of the Distribution of Radiation Dose of Electron Beams Used in Superficial Tumours Treatment. TUJ-BA [Internet]. 2019Feb.28 [cited 2024Nov.24];41(1). Available from: https://journal.tishreen.edu.sy/index.php/bassnc/article/view/8475