In This Study we adopt Prepration a new carbon paste electrode was prepared  for   lanthanum  ions  based on a (5% complex: ( 1,3-di phenyl - 5 - paranitrophenyl Formazan )  With lanthanum Ions as  Ionophore ( 42.5% Dibutyle Phthalate , 0.14% Tetra Phenyl Borate Sodium ,52.5% Graphite Powder) , This electrode revealed  a good selectivity for Lanthanum ions over a wide Linear Range of concentration ( 1.0 ˣ 10-7 – 1.0 ˣ 10-1 ) M . the electrode exhibit a Nernstain slope of (19.54±1.00) mv per decade of lanthanum Ions , the detection limit was ( 1.58 ˣ 10 –8 ) M . The response time was about (20) s and The Electrode can be used for 6 months at least; the electrode can be used in the PH range: (3 – 8) without a considerable divergence potential. By using Matched  potential  Methods ( MPM ) , The Proposed electrode show a very good selectivity towards  La 3 +  ion in  presence of some Transition , Heavy metal ions with the relationship Coefficient ( 0.999 ) ,  it was used to determine lanthanum ions in  potentiometric titration  of  lanthanum ion with  EDTA , and to determine lanthanum ions in  realistic environmental samples.